| Product Performance | |
| Long-Term Stability | ±0.2%FS/year(Not cumulative) |
| Accuracy | ±0.25%FS, ±0.5%FS(Optional) |
| P201 Temp. Effect | 1.5FS/100℃ |
| Comp. Temp. | -20℃~80℃ |
| Working Temp. | |
| Non-Display | -40℃~85℃ |
| LCD | -20℃~70℃ |
| Measuring Medium Temp. | |
| Filled Silicone Oil | -40℃~125℃ |
| -40℃~250C(Adding connectors) | |
| Zero Error | ±1%FS |
| Span Error | ±1%FS |
| Response Time | ≤30ms |
| Fatigue Life | More than 100million designs |
| Physical Characteristics | |
| Pressure Interface | See Ordering Guide |
| P201 Contact Parts | SS316L |
| Electrical Interface | See Ordering Guide |
| Shell Material | SS304 |
| Shell Protection | IP65/IP67 |
| Shock | EC 68-2-27 procedure I |
| Vibration | 10g, 25-2000Hz |
| Weight | About 200g |
| Environmental Data | |
| Certification | CE, RoHS |
| Voltage Output | |
| Output (3-wire) | Min OV to max 10V |
| Power Supply | Over full range output 4V, max 30Vdc |
| Current Output | |
| Output (2-wire) | 4-20mA, 4-20mA+HART 2-wire |
| Power Supply | 10-30Vdc |
| Range | Proof Pressure(xFS) | Burst Pressure(xFS) |
| Kpa/bar | X-100P | X-100P |
| Vacuum | 5.0xFS | 10.0xFS |
| 40~100kpa | 2.0xFS | 5.0xFS |
| 200kpa~35bar | 2.0xFS | 3.0xFS |
| Modbus-RTU | ||||||
| Type | Terminal | Current signal | Voltage signal | Voltage signal | RS485 | Cable Colour |
| M12*1 | 1 | +ve Supply | +ve Supply | +ve Supply | +ve Su pply | Brown |
| 2 | OV Com | -ve Output | -ve Supply | White | ||
| 3 | tve Output | +ve Output | +ve Output | A+ | Blue | |
| 4 | -ve Sup pply | B- | Black | |||
| DIN | 1 | +ve Supply | +ve S upply | +ve Sup pply | +ve Supply | Red |
| 2 | +ve Output | +ve Output | -ve Sup ply | -ve Supply | Black | |
| 3 | OV Com | +ve O utput | A+ | Blue | ||
| GND | -ve Out put | B- | White |

Flat embedded diaphragm type
Model: X-100P-①②③④⑤⑥
| ① | ② | ③ | ④ | ⑤ | ⑥ | |
| Attention | Range | Process connection | Output | Electrical connection | Radiator | Accuracy |
| 1. 4PIN Standardized 2m Cable Line
2. Direct lead with 1 m cable 3. Voltage output is not optional with 8L 4. Can only be selected with 2D8L 5. With»*»,please contact the technician |
Range
K=kPa B=barM=MPa |
2G*=G1/4 *
4G=G1/2 6G=G3/4 8G=G1″» 20M=M20x1.5XX*=Other *» |
4A=4~20mA
4AX=4-20mA+LCD With Display 10V=0~10V |
2D=DIN43650
Hirschmann 4P=4pin xE=Direct lead x meter 8L=2088 Aluminium shell |
=135℃
No radiator 5S=150℃ 5 radiators 10S=200℃ 10 radiators |
=±0.5%FS
2=±0.25%FS |